Serial number

malfunction

the reason

exclude

1

The display is not lit

·

The balance is not properly connected to the power supply

·

Fuse damage

Check the reason for not turning on, reconnect

One can be changed, if it is damaged, it must be sent to the maintenance unit.

2

Weighing display is unstable (data jitter)

·

Worktable instability

·

Airflow

·

The plate is not placed or has objects touched

Choose to place a solid, stable workbench

Check if the windshield is closed

Put the weighing plate and take the touch object on the edge of the plate.

3

Peeling (clearing)

Do not display zero

·

Workbench is not stable, peeling (clearing) operation is too powerful

Choose a sturdy workbench and don't use too much peeling.

4

Weighing result is incorrect

·

The balance is not at zero when empty

·

The balance is not calibrated or the calibration weight used is not allowed

Press the tare key (clear) when empty.

Recalibrate with the correct calibration weight.